"We found a less pronounced amount of contrast material exited the end holes of the modified catheter as compared with the standard end-hole-only catheter," said Rendon C. Nelson, MD, senior author of the study. "Numeric calculations showed that the addition of side holes or slits resulted in a nine to 30 percent reduction of the velocity of contrast material exiting the end hole of the catheter. And we saw more of a cloud like dispersal rather than a jet," said Dr. Nelson.
"The rate at which these injuries occur varies from practice to practice. We typically have 14-15 patients per month who suffer from them. Normally the injuries we see are mild, including pain and swelling at the injection site, however they can be more severe," he said.
"As our study suggests, the development of a modified catheter that decreases the jet like phenomenon seen with a standard-end-hole catheter could improve patient safety and decrease the likelihood of injury," said Dr. Nelson.